CVE-2025-4393
Medtronic MyCareLink Patient Monitor has an internal service that deserializes data, which allows a local attacker to interact with the service by crafting a binary payload to crash the service or elevate privileges. This issue affects MyCareLink Patient Monitor models 24950 and 24952: before June 25, 2025
- Score
- CVSS 3.1 6.5 Medium
- Status
- Deferred - security@medtronic.com
- Published
- 2025-07-24
- Last modified
- 2026-03-27
- Weaknesses
- CWE-502
Affected products
References
- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-205-01
- https://www.medtronic.com/en-us/e/product-security/security-bulletins/mycarelink-patient-monitor-vulnerabilities.html