CVE-2015-3954
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
- Score
- CVSS 3.0 9.8 Critical
- Status
- Modified - ics-cert@hq.dhs.gov
- Published
- 2019-03-25
- Last modified
- 2024-08-06
- Weaknesses
- CWE-285
Affected products
- Plum A3 Infusion System (Hospira)
- Plum A Infusion System (Hospira)
- Symbiq Infusion System (Hospira)
- Plum A+3 Infusion System Firmware (Pifzer)
- Plum A+ Infusion System Firmware (Pifzer)
- Symbiq Infusion System Firmware (Pifzer)
References
- https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01