CVE-2014-7233
GE Healthcare Precision THUNIS-800+ has a default password of (1) 1973 for the factory default System Utilities menu, (2) TH8740 for installation using TH8740_122_Setup.exe, (3) hrml for "Setup and Activation" using DSASetup, and (4) an empty string for Shutter Configuration, which has unspecified impact and attack vectors. NOTE: since these passwords appear to be used to access functionality during installation, this issue might not cross privilege boundaries and might not be a vulnerability.
- Score
- CVSS 2.0 10.0 Critical
- Status
- Modified - cve@mitre.org
- Published
- 2015-08-04
- Last modified
- 2024-08-06
- Weaknesses
- CWE-255
Affected products
References
- http://apps.gehealthcare.com/servlet/ClientServlet/5458232-1EN+r4.pdf?REQ=RAA&DIRECTION=5458232-1EN&FILENAME=5458232-1E…
- http://www.forbes.com/sites/thomasbrewster/2015/07/10/vulnerable-breasts/
- https://ics-cert.us-cert.gov/advisories/ICSMA-18-037-02
- https://twitter.com/digitalbond/status/619250429751222277